Florida Department of Environmental Protection Florida Department of Environmental Protection
 
* DEP Home * About DEP * Programs * Contact * Site Map * Search
MyFlorida.com  
tab corner of content areaAssessment and Restoration Support

Guidance on Choice of Turbidimeters
 

EPA Method 180.1 specifies criteria to use in determining an acceptable turbidimeter. These may be found in Section 5.4 of the method. Standard Method 2130 B specifies criteria to use in determining an acceptable turbidimeter. These may be found in Section 2.a of the method. The criteria specified are common to both methods and are listed below:

Light source: Tungsten lamp operated at a color temperature between 2200-3000 oK.

Distance traversed by incident light and scattered light within the sample tube: Total not to exceed 10 cm.

Detector: Centered at 90o to the incident light path and not to exceed +/- 30o from 90o (e.g. between 60o-120o). The Detector and filter system, if used, shall have a spectral peak response between 400-600 nm.

In the Horiba Water Quality Checker U-10 the scattered light is measured at a point 30o offset from the light source. This does not meet the detector requirements.

In the LaMotte Model 2008 the spectral peak response approached 900 nm. A letter from James O'Dell the ATP Coordinator for EPA states that the LaMotte Model 2008 does meet design criteria and is acceptable for compliance monitoring. Submittal of an Alternate Test Procedure (ATP) application is not required.

Turbidimeters for Groundwater Purge Stabilization Measurements, Oct. 30, 2002 (.pdf, 16kB)

Water Quality Standards Program
Florida Department of Environmental Protection
2600 Blair Stone Road - M.S. 6511
Tallahassee, FL 32399
850-245-8346
Contact Us

Last updated: May 05, 2014

  2600 Blair Stone Road M.S. 3500   Tallahassee, Florida 32399   850-245-8336 (phone) / 850-245-8356 (fax) 
DEP Home | About DEP  | Contact Us | Search |  Site Map